Effect of Co60 Sterilization on Long-Term Stability

Procedure:
Tecoflex® extruded tapes were sterilized with single (2.5 megarads and double 5 megarad total) doses of Gamma radiation. The retention of physical properties of these tapes has been recorded at 3 intervals over a 15-month period (see chart). Testing was done in compliance with ASTM D-638-72 for tensile and elongation properties of plastics.

Results:
As the graphs indicate, even after 5 megarads exposure and 15 months post irradiation, the Tecoflex® family of aliphatic polyurethanes continues to retain its outstanding physical properties.

Tensile Strength after Co60Exposure
EG80A
Co60 Exposure
2 Weeks
7 Weeks
15 Months
2.5 Mrad
-2%
-7%
0%
5 Mrad
-11%
-8%
-1%

Tensile Strength after Co60Exposure

EG60D
Co60 Exposure
2 Weeks
7 Weeks
15 Months
2.5 Mrad
-3%
-8%
-2%
5 Mrad
-15%
-21%
-25%


TOP



Elongation Change after Co60Exposure

EG80A
Co60 Exposure
2 Weeks
7 Weeks
15 Months
2.5 Mrad
+12%
+16%
+7%
5 Mrad
+16%
+17%
+14%

Elongation Change after Co60Exposure

EG60D
Co60 Exposure
2 Weeks
7 Weeks
15 Months
2.5 Mrad
-3%
+13%
0%
5 Mrad
-5%
+5%
-10%